LiQuilaz Particle Counter Maintenance
LiQuilaz Particle Counter Maintenance

This is a basic guide to LiQuilaz and LiQuilaz II particle counter maintenance. Occasional cleaning of the liquid capillary is required and it is the customer’s responsibility to perform the cleaning when necessary.

The need for capillary cleaning and the general health of the capillary can be improved by continuously running water through the system when it is not in use. In addition, never allow the capillary to dry out, as this could permanently damage the capillary.

LiQuilaz Particle Counter Cleaning Procedure Using Syringes

This is a basic guide to LiQuilaz and LiQuilaz II particle counter maintenance. Occasional cleaning of the liquid
capillary is required and it is the customer’s responsibility to perform the cleaning when necessary.

The need for capillary cleaning and the general health of the capillary can be improved by continuously running
water through the system when it is not in use. In addition, never allow the capillary to dry out, as this could
permanently damage the capillary.