Monitoraggio di nanoparticelle in ambienti di produzione ultrapuliti

Abstract

NP Monitoring in Ultra-Clean Mfg EnvironmentsWith its 10 nm sensitivity, the Nano-ID® NPC10 NanoParticle Counter bridges the gap between conventional aerosol optical particle counting at 0.1 μm (100 nm) and sub-nm scale airborne molecular contamination. Experimental and field data demonstrate that ISO Class 1 and 2 environments, which have very low concentrations of particles greater than 100 nm, can possess much higher concentrations of nanoparticles. This suggests that the current ISO classification system may be insufficient to properly describe the cleanliness of the high-purity environments required for advanced electronics manufacturing. This application note presents comparison data from a Nano-ID NPC10 NanoParticle Counter and a traditional optical particle counter across ISO Class 1–4 environments, the challenges associated with testing at these particle concentration levels, and a side-by-side comparison of two Nano-ID NPC10 NanoParticle Counters.

Interested in learning more? Complete the form on this page to receive a free copy of the Nanoparticle monitoring in ultra-clean manufacturing environments application note.

In order to provide complete functionality, this web site needs your explicit consent to store browser cookies. If you don't allow cookies, you may not be able to use certain features of the web site including but not limited to: log in, buy products, see personalized content, switch between site cultures. It is recommended that you allow all cookies.