Planning and Installing an Environmental Particle Monitoring System (489.4 KB)

Download above file to see all tables and figures.
The HTML version below provides an overview. Download the pdf file to see all tables and figures.

Portable Particle Counter Lasair II

Introduction

For aseptic manufacture of pharmaceutical products there has been a shift, primarily due to legislative regulations. Traditionally, particle monitoring has been the classic portable monitoring of a cleanroom. New regulations have lead to a requirement for an automated, remote particle counting solution. This paper reviews the various steps for the implementation of an automated particle monitoring solution for a non-viable particle counting system. The steps also apply if you later enhance the non-viable particle counting system to include viable particle counters or other environmental parameter components.

There are several steps to be followed for the implementation of a particle counter system and the GAMP 3 guidance for the validation of these systems certainly forms the core of the requirements. A typical particle monitoring project follows this format: design, build, install, test, and validate. Each of which has its own time line.

Building the Timeline

The key to a successful project is to ensure that all phases of the implementation are executed in a timely manner. The identification of each of the major steps can be presented as a GANT chart which will also identify when obstacles, such as shut downs, need to be accommodated. Figure 1 shows a typical timeline GANT chart for a complete particle counting project.

Figure 1 Project plan outline as a GANT chart. (Download this paper for all tables and figures) (489.4 KB)

Each of these summary tasks can be broken down into individual tasks so that resources, both material and labor, are available. This becomes more critical when multiple components of a project are encroaching on each other. This often occurs during particle counter installation when different trades are vying for the same space and during particle counter validation when deadlines are tight.

System Design

There are several documents that identify how best to design a particle monitoring system. Considerations include selection of sample points of the particle counters, which hardware to use for each application, and the relationship between risk verses instrumentation. Figure 2 shows a typical system designed to monitor the environmental conditions within a cleanroom.

Figure 2 Facility Monitoring System Schematic (Download this paper for all tables and figures) (489.4 KB)

The driving factors in designing a particle monitoring clean room should be defined in the User Requirement Specification (URS), the contents and format of which is fully described in the GAMP guidance documentation. Any changes required to a system that makes it user friendly should be incorporated as early in the project as possible to avoid costly changes to project scope later. When the speed required to implement a system does not allow sufficient time for the generation of a 'perfect' URS it can be subcontracted to a design company for preparation. It is important to ensure the end users or system owners review the document as the liability of using a system lies with them.

System Installation

Lasair III 100 LPM Particle counter As can be seen from Figure 2 there are many components to the installation of a monitoring system. These include:

  • Wiring: The infrastructure of wiring includes power cables (24VDC, 120-240VAC, and 3 phase), signal wiring 4-20mA analog, Ethernet and RS485 data communications, and I/O connections for process controls, starting stopping of vacuum pumps and the integration of functional hardware switches for state changes, etc. The choice of wire depends on load, fire hazards, length of runs, and signal speed. Cables are normally run in either cable tray or conduit specific for a location, where existing support infrastructure can be incorporated; this reduces redundancy in the installation.
  • Tubing: There are primarily two types of tubing: one that draws a sample to the sensor (sample tubing) and one that enables a sensor to operate (vacuum tubing). Particle counter tubing should not exceed 2 m due to excessive losses of large particles. It is also important to ensure that any bends are maintained at the largest possible radius to reduce losses through transportation.

If the particle counter does not have an integrated vacuum pump and requires an external vacuum source, the pump should be installed at a distance from the particle sensor in a technical space (mezzanine) and the tubing run via a manifold header and a tubing drop to the particle sensor. The design of the particle counter manifold is site specific. It is desired to install a loop manifold as this reduces differences in pressure drop across the system. The limitation of distance and installation criteria for the vacuum system is different than for sample tubing, as airflow is now post measurement. Figure 3 shows a loop manifold design and independent drops down to each individual sensor.

Figure 3 Vacuum Manifold Schematic (Download this paper for all tables and figures) (489.4 KB)

  • Particle Counters: The installation of particle or other sensors is either mounted inside the filling equipment, on the filling equipment with sample tubing to draw a sample to the sensor, on/in the wall of the background room, or mounted on a supply/exhaust duct. The services, wiring, and tubing are run to the sensor via local conduit or service runs through equipment. These services are then taken back to a central location where the vacuum system and data cabinet reside.
  • Data Cabinet and Vacuum Pump: Central to the system is the control data cabinet and vacuum system pump(s). They are typically mounted in the technical space as access for maintenance is required on a routine basis. The size and type of cabinet and vacuum pump depend on the size of the system, number of particle counters, voltage requirements, etc. Figure 4 shows two example data cabinets installed in technical spaces.

Viable Particle Counter: MiniCapt Figure 4 Data Cabinets (Download this paper for all tables and figures) (489.4 KB)

  • Computing and Network: All particle counters report to a centralized monitoring system. Control and reporting occurs here. The location of this system is flexible; all sensors are on an Ethernet network, therefore only a single connection is required to the central computer. If displayed or reported data is required at remote locations, a classic network architecture can be designed. Connections to site networks can be made via system routers, gateways, or other bridging means. A suitable desk in an office is used for the central computer and remote terminals can be installed into the fabric of the cleanroom using suitably compatible computer terminals or HMIs.

System Validation

The validation of the installed system follows the Good Automated Manufacturing Practices Guide. This continues from the initial URS and the design documents. A matrix of key components identified in the URS are linked to suitable tests to prove the robustness of the system, including error reporting and data archiving. This allows a traceable link between the requirements and the finished system, relative to the risk of the system. The testing proves that the installation of all the system components has been done in accordance with standards and as per the detailed design document (IQ) and that the system operates in accordance with the expectations made of it (OQ), i.e. when it exceeds a preconfigured threshold and alarm annunciation is activated.

Validation typically takes a few weeks depending on the size and complexity of the design. Once the validation testing is completed it can be turned over to the end user to begin the process of integration. This is an additional layer of validation to prove that it does what it was intended to do, i.e.: support the release of a batch through proof of control over the manufacturing environment (PQ).

System Implementation

This final stage causes the most concern with many new users of systems as the volume of data generated is much greater than that historically generated using a portable solution. The data can no longer be seen as a single page of data. A single particle counter can generate 5-10 pages of data per day, if this is extrapolated to a 10 sensor system, 50 -100 pages can be reported, so how do you deal with this volume of particle counter data?

System training provides the users with the skills they need. Training is in two parts; part one shows users how to use the system, change alarm limits, create new reports, add and disable user accounts, modify configuration variables (change control permitting of course). Part two of the training occurs after the system has been in use for several months. This second part of training is the key to a successful implementation, reading the data to deliver information relative to proving environmental control, necessary for compliance to the current standards.

Summary

Strong planning, following GAMP guidelines, and proper training to understand how to interpret the system data are essential for an effective particle monitoring program. A successful implementation requires planning, implementation, testing, and training to understand the particle counter results. Particle Measuring Systems provides services for all these steps.

(Download this paper for all tables and figures) (489.4 KB)

Authors

Mark Hallworth Pharmaceutical Business Manager.

Edward Applen Project Services Manager

  1. EC Guide to Good Manufacturing Practice, Revision to ANNEX 1, European Commission, Enterprise Directorate General, September 2003.
  2. EC Guide to Good Manufacturing Practice, Revision to ANNEX 1, European Commission, Enterprise Directorate General, September 2003.
  3. GAMP Guide for validation of Automated Systems; December 2001, published by the ISPE.
  4. Choosing the most suitable non-viable sample point locations, Technical Note 79, M Hallworth, Particle Measuring Systems 2007.
  5. Particle monitoring requirements in Pharmaceutical Cleanrooms, Application Note 41, M Hallworth, Particle Measuring Systems 2005.
  6. An Analysis of Acceptable Particle Losses in Tubing, Application Note 81, M Hallworth, Particle Measuring Systems 2007.

Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801.

Without Measurement There Is No Control