Particle Counting Statistics in DI Water (178.3 KB)

Download above file to see all tables and figures.
The HTML version below provides an overview. Download the pdf file to see all tables and figures.

Particle Counter for DI water: Ultra DI®

Introduction

State of the art DI water systems are now so clean that quantifying the particle concentrations is difficult. When measuring the cleanliness of the water the scarcity of particles can lead to statistical variations that can be as large or larger than the actual particle concentrations. A thorough understanding of the statistics of particle counting is needed to mitigate these variations and effectively measure the particle contamination.

There are three major determinates of the statistical performance of a particle counter. First is the particle size distribution, second is the sample volume, and third is the size sensitivity. In addition, the sample interval of the particle counter used will also affect the variation in the measurements. All of these factors, as well as appropriate alarm levels, must be considered when determining which particle counter to choose.

This paper will discuss the size distribution of particles found in ultrapure water systems, explain the statistics of particle counting, and show how sensitivity and sample volume both affect the statistical performance of particle counters. Finally, a spreadsheet has been developed which compares the statistical performance of different particle counters, shows the variability in particle counters that can be expected, and estimates the number of times a day a given alarm level will be exceeded due to statistical variations.

Particle Size Distributions in Ultrapure Water

Knollenberg and Veal, in Optical Particle Monitors, Counters and Spectrometers: Performance Characterization, Comparison and Use, did an extensive study of particle size distributions in clean water systems. They examined 17 particle counters and the results are shown in Figure 1. The distributions they measured could best be represented by a power law expression of. (Download paper for all figures.) (178.3 KB)

A subsequent study by Mitchell(of Particle Measuring Systems), "Statistical Analysis of Particle Instruments for Liquid-borne Particles: Understanding the Impact of Size Sensitivity and Sample Volume" showed that this power law distribution holds for all liquid systems. Further, the relationship holds true for chemicals, as well as water, that are filtered and in a steady state. Significant deviation from this relationship occurs only when particles are actively being created, such as during etching processes or during ultrasonic cleaning.

Based on this evidence it is reasonable to use the relationship to estimate the number of particles at different sizes. That is, we can assume that if we have 8 particles per ml at 0.05 mm there will be only 1 particle per ml at 0.1 mm.

Basic Particle Counting Statistics

The statistics of liquid borne particle counters are relatively simple and well known. It has been established that particles are counted in time in accordance with a Poisson process with rate l(3). l is the expected rate or the long term average rate, and is often established empirically. The expected number of particles counted for any period of time t is then lt. So if a large number of measurements are made during time period t the average value of these measurements is lt. The standard deviation of these measurements will be the square root of the expected value (4):(see pdf)

Two important observations can be made from equations 1 and 2. First, as more particles are counted, the statistical variation decreases relative to the measurement average. That is, the ratio of the average to the standard deviation, or the signal-to-noise ratio (s/n) on the measurements, increases as more particles are counted. Relative to particle counters this means that the more particles a particle counter counts per unit time the better the statistical performance. Second, longer measurement intervals also increase the signal-to-noise ratio, and the improvement is proportional to the square root of the time increase. If the sample interval is increased by a factor of 4, the s/n will increase by a factor of 2. If the sample interval is increased by a factor of 10 the s/n will improve by a factor of 3.16.

If the long term average is known, equation 3 provides the ability to estimate the number of times a certain measurement value will be seen. This is important because we can now estimate the number of times a particular value will be exceeded and this can be essential in setting appropriate alarm levels and sample intervals.

Statistics and the Real World

An example of the impact of statistical variation is shown in Figure 2. (Download paper for all tables and figures.) (178.3 KB) These measurements were made on the water system at a major semiconductor facility. This facility had a specification of less than 4 counts/ml, the sample volume of the particle counter was 0.25 ml/min, and they were using 4 minute sample intervals. The long-term average of the measurements is 3.2 counts/ml, so the particle counter was meeting specifications, but the statistical variation in the 4 minute measurements caused many of these measurements to exceed their specification. The customer was concerned because there are several data points at 6 counts or higher even though the mean for the day was only 3.2 counts. Based on the mean and using Poisson statistics, we can calculate how many times in a day we would expect to see each number of counts. The results are shown in Table 1 and plotted in Figure 3.

As shown in Table 1, using the Poisson distribution one would expect 6 counts, 20 times over the day, but only 17 actually happened. The results are plotted in Figure 2, and as can be seen the agreement between theory and measurement is very good. Therefore, the occasional high counts seen in Figure 1 are simply the result of the statistical variation, and reacting to these high counts is inappropriate.

Estimating Particle Counter Performance

Assuming a distribution, the user can now compare the statistical performance of different particle counters based on their sample volume and sensitivity. For example, we can compare two volumetric particle counters (instruments that measure 100% of the sample flow) that both have sample volume of 50 ml/min, with particle counter #1 having 0.1 mm sensitivity while particle counter #2 is limited to 0.2 mm. particle counter #1 will see 8 times as many particles in a given period of time based on (diameter)-3 distribution. That is,(see pdf).

Therefore, the standard deviation for particle counter #1 is 2.8 times larger in absolute terms. However, as a fraction of the number of particles counted it is 8 / 2.8 = 2.8 times smaller. This reduces the affect noise has on the data and the user will be able to detect smaller changes in the particle concentration.

Another example of sample volume's effect on data is seen by comparing three particle counters designed specifically to monitor DI water systems. Table 2 shows the relevant specifications of these particle counters, with all three sensitive to 0.05 mm particles, but having different sample volumes. Very clean water systems now have particle levels of 0.2 particles/ ml > 0.05 mm, and for this example we will assume this to be the "true" particle concentration and/or the long term average. A sample interval of 30 minutes, typical for monitoring DI water systems, will be assumed.

As shown in Table 2, the standard deviation decreases as the sample volume increases, which in turn produces fewer alarm conditions as the statistics are dependent on the number of particles detected and not the concentration. Therefore, all calculations are based in particle numbers, with the effect of sample volume on standard deviation being expresses as (see pdf).

As part of this paper a spreadsheet has been developed to perform this type of comparison previously seen in Table 2. The spreadsheet requires the user to input the particle concentration for a user-defined size, the instrument specifications of sample volume and sensitivity, and the sample interval. In addition the user can specify an allowable variation in the measurement or alarm level. The calculations then assume the (diameter)-3 relationship and statistical nature of how the particles arrive in time.

Table 3 shows another comparison between three instruments that are designed to monitor DI water systems. The user first enters in the assumed particle concentration and in this case the concentration is 0.8 particles /ml at 0.05 mm. Next the sample interval is entered in seconds, here 600 seconds or 10 minutes. Finally, an alarm level is entered. The alarm level is relative to the average concentration, and in this case the alarm limit is specified 50% higher than the long-term average, meaning the alarm level is 1.2 counts /ml. Next the sample volume and first channel sensitivity of the particle counters are entered. Remember that the sample volume is sometimes not specified directly and may be calculated by multiplying the sample flow rate by the percent sampled.

In this example all three particle counters have the same sensitivity and differ only in the sample volume. Particle counter 3 outperforms the other two particle counters because of its relatively large sample volume as seen in the number of particles detected per measurement, the standard deviation, and the number of alarms per day. The number of alarms per day is particularly important because this demonstrates how better statistical performance affects the ability of a particle counter to monitor the water system. With particle counter 3 a slight increase in concentration will be detected, while the statistical variation on the other two particle counters creates 54 false alarms for Instrument 1 and about 34 for Instrument 2. To limit the number of false alarms to less than 1 a day, the alarm level must be close to 389% of the average concentration or 3.91 particles /ml for Instrument 1 and 174% or 2.19 particles /ml for Instrument 2. Increases in particle levels must be relatively large or will be lost in the noise for Instruments 1 and 2. On the other hand, Instrument 3 with the larger sample volume will be able to recognize small increases in concentration.

Many users have requested particle counters with sensitivity to smaller particles, but the best commercially available instrument is sensitive to 0.05 mm. To examine the benefit of increased sensitivity the user can specify different sensitivities and sample volumes as shown in Table 4. The particle concentration is assumed to be 0.4/ml at 0.05 mm and the alarm 50% above the average.

Summary

The statistical behavior of particles in ultrapure water systems can be understood and predicted using Poisson statistics. Combined with the (diameter)-3 size distribution demonstrated by Knollenberg and Veal and Mitchell the statistical performance of particle counters can be estimated and comparisons of different particle counters can be made. These comparisons show that often sample volume is more important than increased sensitivity.

As can be seen, the particle counter with the best sensitivity, Instrument 1, has the poorest statistical performance, and the instrument with the poorest size sensitivity, Instrument 3, has the best statistical performance. Under the assumed conditions, Instrument 1 has approximately 20 alarms because of statistical variation while Instrument 3 has less than one. In the end, the most important factor determining statistical performance is the total number of particles measured in a sample interval, not the size of the particles. A particle counter's sample volume and sensitivity must be considered together to estimate particle counter performance.

(Download paper with all tables and figures.) (178.3 KB)

Ultra DI® is a registered trademark of Particle Measuring Systems, Inc.

Contact us if you need more information or have questions.

Last Updated: 2000

Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801.

For more information go to Ultra DI UPW particle monitor.

Without Measurement There Is No Control