Basic Guide to Particle Counters and Particle Counting (8.4 MB)(Download this paper for all tables and figures) (8.4 MB) Particles Sources Particles can be produced by many different sources. Inert (nonliving) particles usually arise from the rubbing of one item against another, such as the dust produced when you saw through a piece of wood. Humans shed lots of inert particles, as in the continuous sloughing off of dead skin. Electric motors give off particles where the commutator is rubbed by wire brushes. Plastic disintegrating slowly in ultraviolet light sheds particles in a light breeze. Viable particles are living microorganisms such as bacteria, viruses and fungi. Humans shed large quantities of viable particles. Particles can be classified as organic (arising from living matter, though not necessarily alive themselves) or inorganic (arising from matter that was never alive). A dead skin cell is an inert organic particle. A protozoan is a viable organic particle. A grain of copper dust is an inert inorganic particle. Particle Sizes The next larger family of particles is atoms. Next in size are molecules, or groups of atoms. These are still too small to be considered industrial microcontamination. The particles that our particle counters monitor for clean manufacturing range in size from well under a micron (abbreviated µm, 1/1000 of a millimeter) to about 100 microns (1/10 of a millimeter.) Particles larger than this can be seen with the naked eye. Particles smaller than this (approximately 0.01 µm) are of little or no consequence to modern manufacturing processes, as was noted above. There are several different ways to measure the size of a particle. The diagram below (download pdf for all diagrams and tables) shows the standard dimensions that are employed to describe a particle. A sphere is shown in dashed lines representing a PSL (polystyrene latex) particle, which is a man-made particle used to calibrate particle counters and test filters. The scientific term for each type of measurement is shown above. Each is useful in different contexts, especially in microscopy. Some particles can change in size. Take, for example, a viable organic particle like a paramecium. A paramecium is a microorganism that, like most animals, is made mostly of water. If the paramecium becomes desiccated (dries up) it will be much smaller than it was when it was hydrated (full of water). For more information on particle size, see the box on the next page. Why is the size of a particle of interest to a manufacturer? Depending on the clean process, particles in a particular size range may be of interest because they could do specific kinds of damage to the process. If you are buying a filter, you need to know how small the pores in the filter media need to be. Material Particles can be made of any substance and can be organic or inorganic in origin. Metals, plastics, fibers, animals, sea salt, smoke, fumes, and dust are all examples of particle sources. Virtually anything can generate particles under the right circumstances. In a cleanroom, the most prolific particle generators are usually the people who work inside, shedding skin cells, breathing, sneezing, etc. To put these items in perspective, a typical human hair is about 50 -150 µm in diameter. (Download this paper for all tables and figures) (8.4 MB) Behavior: How do particles act? Movement Currents: Currents are the laminar (smooth) and turbulent (rough) movements of a fluid. Currents are a result of pressure differences, with the fluid always moving from an area of higher pressure to an area of lower pressure. Particles suspended in a laminar flow tend to remain in that part of the fluid. In air, a lateral (side-to-side) movement is called advection; a vertical (up and down) movement is called convection. Thermal variation (thermophoresis): Temperature differences in a fluid contribute to currents, particularly convective (vertical) currents. Warming a fluid will also increase Brownian motion. This causes the molecules to be more energetic, and consequently they collide more frequently and are farther apart. This is why warm air is less dense than cold air and tends to rise. Brownian motion: Air is chock-full of particles, ranging from visible dust to non-visible gas molecules, that are continually colliding and bouncing off of each other (and into other particles). The same thing is true of liquids. Over time, Brownian motion results in a more-or-less random distribution of particles. The distance a particle can travel in a straight line before it bounces off another particle is its mean free path. Ballistic Forces: Adhesion Electrostatic adhesion: Particles can carry static electricity the same way a balloon rubbed against your hair can. This causes particles to be attracted and stick to a surface that carries the opposite charge. Agglomeration: In liquids, particles tend to agglomerate around (stick to) gas bubbles. Accretion: Particles can stick to each other. This can be the result of electrostatic adhesion or other "sticky" forces. Under certain conditions, it is common for two particles to stick together forming a doublet. Friction: A particle can get caught on a rough surface where the movement forces are not strong enough to dislodge it. This mechanism, along with electrostatic adhesion, is the basis for most types of filtration. Movement and Adhesion Cycle Diffusion and adhesion coexist in a continuous cycle, such that a particle circulates, is trapped, breaks free, circulates, etc. Because of this the number and size of particles in a given fluid volume is always changing. This is important, as we will see later in our discussion of particle counters. (Download this paper for all tables and figures) (8.4 MB) Section II: Environments Many of our modern, high-technology clean manufacturing practices demand cleanliness. Specifically, they demand an absence of particle contamination. To ensure the area is clean, particle counters are required. Let's look at a concrete example, semiconductors, commonly referred to as "microchips." A microchip is a flat piece of silicon on which very small traces (flat wires) are etched, forming transistors and other components. This allows the manufacturer to create a very tiny electronic circuit. Some traces are so close together (0.3 µm apart) that a particle lying across two of them would cause a short circuit. Because of this, the manufacturer wants to filter out any particles in the air that are 0.3 µm or larger. Particles smaller than this are not big enough to cause a short circuit. A particle counter is needed to ensure the particles were filtered properly so that the product is protected. Also, microchips are multi-layered devices, each layer being extremely thin. The result of this is that, for manufacturing purposes, a microchip's effective surface area is equal to the length times the width times the number of layers, compounding the likelihood that a stray particle could shut the whole thing down. The semiconductor maker has to manage the production environment, including with particle counters, to eliminate particle contamination. Another example may be drawn from the pharmaceutical industry: An injectible drug must be free of particles that could block a blood vessel, causing a stroke (interruption of blood supply to part of the brain) or necrosis (interruption of blood supply to tissue). The drug maker, like the microchip maker, has to manage the production environment to eliminate particle contamination, then use particle counters to ensure the system is working correctly. Controlling Particle Contamination There are three ways to control particles:
Filtration Particles are eliminated by filtration. In filtration, the particle-containing medium is passed through a filter with holes large enough for the medium's molecules to pass, yet too small to allow the particles through. A filter accumulates particles during its service life, and is typically replaced before it is saturated (completely full of particles), though some filter media can be purged (cleaned) and reused. There are two steps to filtration; directing the particles to the filter and catching them in the filter. Directing particles to the filter is the more difficult part of filtration. In Section One we discussed the factors involved in particle migration; now think about them in the context of a typical manufacturing facility. Such a facility has an enormous number of particle traps, a great deal of surface area and many sources of contamination. The optimal method of particle management is to preserve laminar flow wherever possible, in the hope that as many particles as possible can be swept through the filter. However, it is not always possible to preserve laminar flow. (Download this paper for all tables and figures) (8.4 MB) Filter media have become very sophisticated and are made of synthetic fibers, porous plastics or ceramics. There are two air filtration standards in current use: HEPA filtration (High Efficiency Particulate Air) is the industry standard for ultraclean or ultrapure manufacturing environments. HEPA filters typically remove 99.99% of particles that are equal or greater in size to the filter specification; usually 0.3 µm. HEPA filtration is often an integral part of an HVAC (Heating/Ventilation/Air Conditioning) system. ULPA filtration (Ultra Efficiency Particulate Air) removes 99.9995% of the particles that are equal or greater in size to 0.12 µm, and is used where ultraclean process environments are required. Particle counters immediately advise when particles slip through the filters. Previously, the filter would be examined with a microscope to count and measure particles. Today, this activity is often performed by particle counters. Particle counting will be discussed later in this guide. Cleanrooms Modern "clean" process environments often must be so clean that it is impractical to simply filter the air in the factory. Separate environments called cleanrooms have been developed to keep particle contamination at known, controlled levels. Cleanrooms are designed to maximize laminar flow and minimize particle traps. In the most efficient facilities, filters are installed in the ceiling and air returns are installed in the floor. This results in the cleanest environment possible, and minimizes the number of particles that are carried from one place to another by advection. Other cleanroom techniques include the wearing of protective gowns, caps, overshoes, and gloves. In the cleanest environments, personnel wear space-suit-like outfits. Cleanroom apparel is an important step in microcontamination control, because people are prolific particle generators. However, it is impossible to fully protect against the possibility of particle contamination, and so particle counters are also required to ensure clean manufacturing. Minienvironments A recent trend has been to use minienvironments, which are in essence miniature cleanrooms with internal robotic arms or integral rubber gloves. Minienvironments are significantly less expensive than cleanrooms, and their use is growing. In some cases it is possible to install a minienvironment in a lower grade cleanroom instead of building a new facility, saving a great deal of money. However, even here, particle contamination can occur and particle counters are needed to protect expensive product. Classification of Cleanrooms and Minienvironments Federal Standard 209 establishes standard classes of air cleanliness for cleanrooms and clean zones based on specified concentrations of airborne particles. A "Class 100" cleanroom has no more than 100 particles larger than 0.5 µm in any given cubic foot of air. A "Class 10" cleanroom would be ten times cleaner than a Class 100 cleanroom. (Download this paper for all tables and figures) (8.4 MB) Standards for Cleanrooms In 1984 the Institute of Environmental Science and Technology drafted IES-RP-CC-006-84-T, "Testing Clean Rooms" a scheme for cleanroom evaluation. The items included are measurement techniques for:
The National Environment Balancing Bureau (NEBB) expanded this standard and offers a third-party certification program. While the program provides useful information, NEBB is not required for cleanroom certification. Cleanroom Evaluation and Certification Cleanrooms are certified after construction or significant physical changes. Certification guarantees that the facility has met the requirements for a statistically valid maximum concentration of specified-size airborne particles. A cleanroom can be certified in any of three different stages: As built: A cleanroom certified "Class X as built" means that the classification was established with the cleanroom fully constructed and ready to use, but without equipment or personnel in the room. At rest: A cleanroom certified "Class X at rest" means that the classification was established with the equipment installed and operating (or operable, as specified) but without personnel within the facility. Operational: A cleanroom certified "Class X operational" means that the classification was established with the room in normal operation, with equipment and personnel performing their normal functions within the facility. Deposited Particles Cleanroom certification schemes generally do not require that the surface deposition of particles be monitored with particle counters-they only evaluate free air. This is a critical issue, because deposited particles may have the greatest impact on manufacturing processes. In order to verify particle deposition, it is necessary to collect deposited particles on a witness plate; a clean, flat object that is known to be free of particles, and is similar in material and characteristics to the item upon which particles are deposited (for example, if you make ABS plastic products, you should use ABS plastic witness plates). Several witness plates are placed throughout the cleanroom, and after a set period of time are collected and the deposited particles are counted. The counting may be done by means of optical microscopy or surface analysis particle counter. Section III: Particle Counters The certification of a cleanroom is an ongoing process. The air quality must be monitored to ensure that the filtration system is working properly and that no unknown particle generators exist. In the early days of clean manufacturing processes, test filters were examined with a microscope to determine the number and size of the particles that were being removed from the air. Sometimes the person counting the particles could tell what the particles were made of (e.g., copper dust). Microscopy is still the best way to learn certain things about contamination, but is limited by the fact that it is a postcontamination assessment technique. In the mid-1950s, the first particle counters were invented. These devices made it possible to monitor particle contamination levels during the manufacturing process, allowing quick responses when contamination levels got too high. This increased productivity. Optical Particle Counters Most people are familiar with the sight of dust in a sunbeam. Four things are necessary for this: sunlight (to illuminate the dust), dust (to reflect the sunlight), air (to carry the dust), and your eye (to see the dust, or more specifically to see the light reflected by the dust). An optical particle counter (OPC) uses the same principles, but refines them to maximize its effectiveness. In modern particle counters, a laser light source is used, the viewing volume is controlled, and a high-sensitivity photodetector is employed to detect light that the particle scatters. Theory of Particle Counter Operation How a Typical Laser Optical Particle Counter Works A generic laser optical particle counter consists of five major systems:
Observations Regarding Laser Particle Counters Particle Counter Observation #1 Particle counters do not directly count particles. They count flashes of light scattered by particles (or shadows cast by backlit particles). This is important, because the amount of light a specific particle scatters or eclipses can vary with several factors, including: The shape of the particle: Particles in the real world are seldom the smooth, spherical latex particles used to calibrate a particle counter. If that particle floats through the viewing volume sideways, it will scatter a different amount of light than if it went through lengthwise. The albedo (shininess) of the particle: Some particles are more reflective than others and cause more light to be scattered onto the photodetector. The photodetector produces a larger pulse, and the particle counter thinks it is a bigger particle than it really is. Conversely, some particles are less reflective and trick the particle counter into thinking a smaller particle has passed through the viewing volume. Particle Counter Observation #2 Particle counters don't count every particle in the room. In fact, depending on the room size, a particle counter in one minute of counting will look at only 1/60,000 of the total air in a five thousand square foot cleanroom with a twelve-foot high ceiling. That's only 0.0000166%. In an hour, it will count sixty times that, or 0.001% of the total volume of air/gas/liquid. Because of this, particle counters must be used to provide a statistically valid sample of the air in a cleanroom (or the liquid in a tank, or the gas in a cylinder.) A statistically valid air sample is a sample that is representative of the average air in the rest of the room. Although this sounds simple, problems can arise since particles are never truly diffused (evenly distributed throughout the container, whether the container is a room or a bottle). They tend to stay in laminar flow, they tend to accumulate inside turbulent flow, they tend to stick to surfaces, and they tend to rise in warm air and hang around near the ceiling. Cleanrooms are designed to minimize these particle traps, but they can seldom be eliminated completely. Types of Particle Counters There are several varieties of particle counters currently available. The primary differences are related to the medium in which particles are suspended: air, liquid, gas, vacuum, or surface. The most common particle counters are used to measure contamination in a HEPA-filtered cleanroom. Liquid particle counters are used for everything from drinking water to injectible drugs to transmission fluid to hydrofluoric acid. Some liquid particle counters require an accessory called a sampler, which draws a precise volume of liquid and passes it through the particle counter at a specified rate; other particle counters are simply attached to a pressurized source. Some particle counters are designed to count particles suspended in a gas. Gases can be inert or volatile and may be dry (anhydrous) or have water vapor suspended in them. Vacuum Particle Counters Some activities performed during semiconductor fabrication occur under vacuum. Particles generated by the fabrication tools and materials are measured with specially-designed particle counters. Surface Particle Counters Semiconductors have to be free of surface contamination before the next layer can be added. Optical components, such as lenses and mirrors, have similar requirements. Surface analysis particle counters use lasers to create a kind of topological map of a surface, showing the location, size and shape of particles. Atmospheric/Meteorological Particle Counters Particle counters are used to examine atmospheric contamination, in applications like pollution control or weather studies. Some of these instruments measure water droplets, ice crystals or condensation nuclei. Variant Technologies of Particle Counters There are several technological variations that can be used in the design of a particle counter. These are dictated by the kind of particle counter, and the task to which it will be applied. Scattering vs. Extinction Volumetric Particle Counters vs. Inline Particle Counters Particle Counter vs. Particle Spectrometer Choosing Between Particle Counters and Particle Counting Spectrometers: The intensity of a laser beam is distributed in a bell curve or Gaussian distribution, i.e., it is brightest in the center. (A discussion of Gaussian distribution is presented later in this guide.) Particles that pass through a laser beam do not feel any obligation to keep to the center of the beam. In the ideal world, when a particle passes through a laser beam, the flash of light is dim at first when the particle enters the edge of the beam, then brightens as the particle crosses the beam's center, and then dims again as the particle passes through the far edge of the beam and exits the viewing volume. In the real world, however, particles are just as likely to transit the edge of the beam, resulting in a dim flash. Unless the "viewed" portion of the laser - the part that is visible to the photodetector-is limited to the center of the beam, it is impossible for the pulse height analyzer to determine if a small electrical pulse was a dim flash caused by a small particle transiting the bright center of the beam or a large particle that only crossed the edge of the beam. Thus, the ability of the particle counter to accurately measure the size of a particle is limited. Particle spectrometers use focusing or masking techniques to limit the viewed portion to only the center of the beam. A particle counter spectrometer has a smaller sample volume and flow rate, but provides specific particle sizing data. A particle spectrometer is the preferred instrument to analyze problems or conduct studies such as filter element testing with a monodispersed particle challenge. There are plenty of applications where the precise size of a specific particle is of no consequence. For those applications, a particle counter (which views the complete beam) is appropriate. For any given particle size sensitivity, a particle counter has a higher flow rate and sample volume. It is the preferred instrument for multi-point monitoring of a deionized (DI) water system or plant and its associated piping system. Condensation Particle Counters (CPC) and Nonvolatile Residue Particle Monitors (NRM) All automated particle counting techniques are minimum-sensitivity-limited. The optical particle counters discussed thus far cannot detect particles that are smaller than some minimum diameter at which the amount of light scattered by a particle generates a signal that is smaller than the background noise level of the "dark" signal (when no particles transit the laser beam). There is a method wherein the particles are "grown" to a larger diameter that is sufficient to be detected by the particle counter. A condensation particle counter (CPC) contains a reservoir of volatile liquid such as butyl alcohol. The sample air flows through a warm chamber where alcohol vapor mixes with the sample air. Next, the sample air and alcohol vapor flow through a cold condensing chamber, and the alcohol vapor becomes super-saturated and condenses upon the particles. Particles as small as .01 µm will be surrounded by a microscopic droplet of alcohol; typically, all the particle/alcohol droplets will end up at about 1-2 µm - a size that is easily detected. The condensation particle counter must be designed so that all excess alcohol diffuses against the walls of the condensing chamber rather than becoming droplets that would add to the particle counts. As with optical particle counters, CPCs with very small minimum detectable particle diameters are more complex and require more maintenance. One may think that it is always better to count the smallest particles possible, so why not always use condensation particle counters? There are some disadvantages to a condensation particle counter versus an optical particle counter. Someone must periodically refill the alcohol reservoirs. Butyl alcohol has an unpleasant odor and non-butyl alcohol condensation particle counters use a fluorocarbon liquid that is expensive. A tipped-over condensation particle counter will spill and have no data output until the flooded parts return to normal. In many environments (Class 1,000 or dirtier), a condensation particle counter would detect so many particles that it could not count fast enough, and the data would be quite erroneous. Also, unlike an optical particle counter, a condensation particle counter cannot report particle size information. Since all particles will grow to the same diameter when using a condensation particle counter, it can only report that a particle was detected-it cannot determine what size the detected particle was. NRMs: Liquids and CPC Of course, as with other forms of optical particle counting an NRM cannot report the chemical makeup of the detected material. Nevertheless, with proper trend analysis NRM data can be used to alert operators to rising amounts of contamination that cannot otherwise be detected without extraordinarily expensive and slow laboratory tests such as atomic absorption spectrometry. NRMs can be extremely useful in, for example, a DI water plant. Huge savings can be realized by not prematurely replacing filter elements. Carefully scheduling DI bed back-washes can also help reduce product production downtime. Using Particle Counters Particle counters are not like most of the equipment on your bench. Because a particle counter looks a little bit like an oscilloscope, people tend to treat them like oscilloscopes. But particle counters just are not as simple as most other electronic devices, and they are much more sensitive to environmental stresses like vibration, RMI (radiomagnetic interference), extremes of heat and cold, and dirt. Particle counters are high-performance electronic instruments. Guidelines for Handling Particle Counters Particle Counter Unpacking: Particle Counter Installation: Particle Counter Storage: Store your particle counter at room temperature (around 70°F/21°C) on a sturdy shelf in a vibration-free environment where it is not likely to be inadvertently damaged by people moving it to get at something else. Unit file: Particle counters need routine maintenance. A typical item is cleaning the optical surfaces of the particle counter. Over time, optical surfaces accumulate dirt that can scatter laser light. This can result in diminished particle counter sensitivity and/or false particle counting. To avoid this, follow the instructions that came with your particle counter. On most particle counters, cleaning must be performed by the user. Be very careful to follow the directions exactly. If you are unsure of what you are doing, do not proceed. Contact the particle counter manufacturer for further instructions. Applications of Particle Counters This section describes how to use particle counting equipment and accessories to collect useful information about particles. Before we begin, there are a few concepts common to all particle counters that are useful to know. Particle Counting and Trend Tracking There are more sophisticated applications for particle counters. These will be discussed later, under the heading for each type of particle counter. Statistically Valid Particle Sample Particle Counter Data Normalization Raw counts: The total number of particles counted in a particular size channel is called a raw count. Raw counts do not relate particle counts to sample volume, and thus do not provide a picture of how dirty the sample medium is. This data is useful in some applications, as well as in calibrating the particle counter. Normalized counts: Normalization is the formatting of data to make it useful by giving it context. In particle counting, the total number of particles counted by the particle counter divided by the sample volume is a normalized count. Bell-Curve Distribution (Gaussian distribution) Particles in the real world seldom fall neatly into 0.30 µm size bins. Most of what particle counters call 0.3 µm particles, in fact, are either a little bigger or a little smaller than 0.3 µm. We call 0.3 µm the nominal size of the particle because it is convenient (instead of calling them, for example, "0.2547µm to 0.3582µm size particles"). The amount that a particle is different from the nominal size is the variance. (And the variance is the square of the standard deviation.) If you were to precisely measure a number of particles at a nominal size of 0.3 µm and then were to graph the results, the graph would look like this: (download complete pdf for all graphs and figures) On this graph, most of the particles counted are pretty close to 0.3 µm. Particles slightly larger or smaller than 0.3 µm are fewer in number. Particles that are much larger or smaller than 0.3 µm are fewest in number. In a particle counter, particles that fall off the curve end up in the next higher or lower size bin. This is called a bell-curve (or Gaussian, or normal) distribution. Section IV: Particle Counter Hardware and Accessories Aerosol Particle Counters Airborne particle counter are used to detect and measure particle contamination in air. Their typical application is to observe particle contamination trends in clean environments, such as cleanrooms or microenvironments. In addition to cleanroom air monitoring, aerosol particle counters can be used to monitor airborne particles inside a large tool. Another common application is filter efficiency monitoring, where air is sampled immediately before entering and immediately after exiting the filter. The number of influent counts (air going into the filter) is compared to the number of effluent counts (air coming out of the filter). The particle counter can be set up such that a sudden decrease in filter efficiency triggers an alarm. As with all particle counters, aerosol particle counters are most effective when they are used to detect changes in contamination trends. Channel sizes for aerosol particle counters range from 0.05 µm at the smallest to several hundred microns at the largest. The number of channels and the size range of each channel may be factory preset or software-controllable, depending on the make and model of particle counter. One typical application of an aerosol particle counter is to place it on a shelf in a cleanroom and set it to trigger an audible alarm if the level of contamination rises above a preset level. Recall that to take a statistically valid sample, you should sample several points around the room. This can be done by:
Aerosol Particle Counter Manifolds Isokinetic Probes In order to take an accurate sample, an isokinetic probe is used at the end of a sample tube on the particle counter. The isokinetic probe captures a sample from moving air (or any fluid) at the same velocity the air is moving. This allows an accurate normalized particle count to be made. Probes that do not capture a fluid while preserving free velocity are said to be anisokinetic. High Pressure Diffuser Environmental Probe Portable Aerosol Counter Portable particle counters are used to pin-point contamination sources in a cleanroom. They use an isokinetic probe at the end of a hose, and often emit different-pitch tones (like a Geiger counter or metal detector) corresponding to different particle concentrations. Liquid Particle Counters Liquid particle counters are used to count particles in almost every kind of liquid, from water to hydrofluoric acid to petrochemicals to injectible drugs. They are often used to monitor filter efficiency or as quality control devices in batch sampling applications. Liquid Particle Samplers Correctly used liquid samplers can prevent cavitation or the creation of bubbles. Bubbles are a problem because they can accumulate particles (agglomeration), and because they can be erroneously counted as particles themselves. Liquid samplers can also reduce or eliminate effervescence (bubbling) by compressing the bubbles out of a liquid. Viewing Modules Corrosives and Plumbing With Particle Counters Particle Measuring Systems uses several different optical materials and plastics for the wetted surfaces of liquid particle counters: Particle Counter Optics: Sapphire: compatible with most chemicals used in the semiconductor industry, including hydrofluoric acid. Magnesium Fluoride: compatible with most chemicals except ammonium fluoride and hydrogen peroxide. Particle Counter Plumbing: PFA Teflon: a plastic used in some sample cells. PFA Teflon is porous to some chemicals. Other materials include Teflon, KalRez,(an extremely expensive O-ring material) and Kel-F. Particle Counter Chemical Compatibility Gas Particle Counters Acquiring and analyzing representative samples of gas can be difficult. Common sampling challenges in semiconductor factories include hooking-up the particle counter to the gas supply. Typically, a large plant outside the chip fabrication (fab) facility purifies nitrogen and pumps it to the fab through large-diameter stainless steel pipes. There are not many particles in semiconductor gas, and the effects of gravity and diffusion can make it hard to capture statistically-valid samples of the few particles that are present. Particle Measuring Systems field personnel have often encountered gas analysis systems consisting of a homemade high pressure diffuser connected to an aerosol particle counter. High pressure diffusers, for particle counters, manufactured by Particle Measuring Systems contain many proprietary features that have evolved after prolonged efforts to analyze ultra-clean gas. Almost none of the homemade diffusers work at all. Their failure to zero-count when sampling filtered gas, and the randomness of the particle counts are all problems Particle Measuring Systems designers had to overcome before high pressure diffusers were ready to market for particle counters. When deciding whether to use a high-pressure gas particle counter or an aerosol particle counter with high-pressure diffuser (HPD), consider the cost of the gas (HPDs consume more gas than is analyzed), the desired sample flow rate, instrument size, particle size/instrument sensitivity and local data display options (consult with your particle instrument sales representative). Vacuum Particle Counters Viewing Modules (Download this paper for all tables and figures) (8.4 MB) Vacuum particle counters provide the users of "process tools" with valuable data that allows them to plan cleaning of the tools, optimize certain tool operations, and possibly reduce the cost of operating the tools by limiting the number of disposable "witness wafers" used. Surface Particle Analysis System Surface particle counters are designed to measure particle contamination when particles are deposited on a flat surface, such as a silicon wafer substrate, high-precision mirror, precision optical coating, liquid crystal display or witness plate. Surface particle counters are the most demonstrably accurate of all particle counters, because they are capable of counting the same number of particles on a test wafer repeatedly. It is impossible to conduct this test with any other particle counter. A typical surface particle counter system consists of the particle counter, a robot and a clean environment. The robot takes a component from the manufacturing process and places it in the surface analysis particle counter. The particle counter scans the component and creates a sort of topographical map of the surface, showing the size, shape and number of particles deposited on the surface. The robot then removes the component from the particle counter and places it on the appropriate shelf, which depends on the amount and location of particles that were detected. Section V: Particle Counter Data Integration Facility Monitoring Systems (FMS) A facility monitoring system is used to allow all of the particle counters, samplers, manifolds, environmental sensors and other microcontamination assessment equipment to communicate with each other and with a central monitoring station. This allows the collection and analysis of particle data, and the correlation of particle counts with events, like a door or valve opening, or a filter failure or a flow reversal in a duct. FMS is usually controlled by a computer with special software that allows the operator to observe the system in action. The computer can be configured to trigger alarms, generate reports and analyze data. A sample manufacturing facility with cleanrooms, particle counters and FMS is shown here. An explanation follows. A. FMS Computer and Software Each particle counter is connected to the FMS computer. The computer serves as a central control station for the particle counters and manifolds, and is a clearinghouse for the data that is collected. The FMS computer can analyze particle data, track particle trends, and trigger local and/or remote alarms when preset conditions are met, such as maximum normalized particle counts, minimum temperature, maximum relative humidity, etc. Data can be displayed by remote computers that are connected to the FMS computer. Aerosol Particle Counter Manifold and Particle Counter: This setup, along with isokinetic probes, is used to monitor particles in Class 10 and Class 100 cleanrooms along with the Class 1000 equipment space area. An aerosol manifold can be an economical way to monitor many different areas, or to monitor several points in the same area to ensure a statistically valid sample. Aerosol particle counter manifolds exhibit a certain amount of particle loss and inter-sample delay, and so are not suitable for all applications. Mixed air sampling can be used to continuously monitor the combined samples before they are sent to the large vacuum pump. This helps overcome the problem of multiplexed ("muxed") sampling; the delay between samples does not cause brief events to be missed. B. Class 100 Cleanroom Certain assembly, test and packaging operations are conducted in this environment, which is monitored by the aerosol particle counter and aerosol particle counter manifold associated with the hardware setup shown at location A. C. Class 10 Cleanroom Certain assembly, test and packaging operations require greater cleanliness than others. These highly-sensitive operations are conducted in this cleanroom. Particle monitoring is performed by the hardware setup shown at location A. D. Remote Particle Counter This device allows people outside the cleanroom to view particle counts. E. Etching Acid Bath In this part of the facility, an acid bath is used to etch the surface of the product. This acid and its disposal are very expensive, and so the acid is filtered and reused as much as possible. Because of this, the facility manager has chosen to use two liquid particle counters and a corrosive liquid sampler to measure filter efficiency. The amount of particle contamination in the acid prior to filtration is compared to the amount of contamination after filtration, resulting in a measurement of filter efficiency. This helps determine when the filter needs to be replaced and also serves as an alarm if a hole should develop in the filter. It also can sound an alarm if the acid is too dirty to use. F. HEPA Filter Efficiency Spot Check An aerosol particle counter with an isokinetic probe has been mounted on a cart and is used to spot-check HEPA filter efficiency in the facility. Every filter is checked as part of a periodic inspection schedule. Transporting Particles through Tubing to the Particle Counter
Because of this, it is useful to be familiar with the following information: Inside Diameter is the diameter of the inside of a round tube. Reynolds Number is a composite figure that takes into account the shape of the tube, the viscosity of the fluid, the smoothness of the inside of the tube, the straightness of the tube, ambient air pressure, temperature and other factors that affect flow rate inside a tube. Pressure Loss is the amount that air pressure decreases the farther the air has to travel in the tube to the particle counter. Thus, if air is pumped into a 7 mm tube at 10 psi, the air pressure at the other end of a 20 meter long tube is 8.6 psi. Gas Velocity is the speed at which gas travels through the line. Particle Loss The following graph (download the pdf at the top of this page for all graphs and images) depicts particle loss in semi-conductive polyester tubing, 3/8" inside diameter, using a Particle Measuring Systems' 3 CFM aerosol particle counter manifold flow rate: To minimize particle line loss, there are preferred aerosol transport line materials (in order of preference, based on particle loss):
Particle Characterization and Sizing (Download this paper for all tables and figures) (8.4 MB) Contact us if you need more information or have questions about particle counters or particle monitoring. Learn more about our microcontamination monitoring services. Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801. For more information go to a comparison chart of all available particle counters. |
|