Is it possible possible for the Lasair II (550) to monitor/measure particulates in a low pressure nitrogen line (50 psi)? If so what accessories would be required to accomplish this?
回答

Non-volatile gas lines are monitored with the Lasair II in two ways. The first is with a High Pressure Diffuser (HPD) and the second is simply to plumb a pressure release 'T' sample port. I have attached an application note that speaks to each method. If you choose to go the High Pressure Diffuser route we offer a 50 LPM HPD for US$2000.