Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute? - 回答
All particle counters must have fluid flow at their rated calibration flow rate. Passing fluid through the particle counter at flow rates exceeding the calibrated flow will cause the instrument to report particle size and concentration inaccurately. There are currently no particle counters on the market that have a flow rate in excess of 1-liter/minute; however this does not limit the ability to measure in high volume applications. You can simply tee off the main line and connect this lower flow rate to the particle counter. It is usually necessary to provide some type of flow controller down stream of the particle counter to regulate the flow within the calibration limits of the sensor. You can then apply the normalized counts/ml value reported by the particle counter to the entire flow stream.
- カスタマー サポート
- 技術サポート
- よくある質問
- 気中微粒子
- 液中微粒子モニタリング
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Flow rate
- PSLs
- Plumb
- Post Service
- Stable counts
- Statistically valid
- 安定するまでの時間
- 粒子分布
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- LaminarCabinet
- Counts/mL Reading
- APSS 200 parameters
- Particles per mL
- Liquistat Nitrogen needed
- HSLIS Pressure
- 製薬/生命科学(ライフサイエンス)における微粒子測定
- パーティクルカウンタサービス
- AMC分子状ケミカル汚染対策について
- ガス分析
- 全般
- ニュースレター
- 技術資料
- パーティクル カウンタ
- 分子モニタ
- サービス
- レンタル&リース
- ニュース
当ウェブサイト上の情報は予告無く変更される場合があります。