We can not get a zero count reading for our APSS 200; the analysis is carried out under laminar class I cabinet. What could be the reason? - 回答
This is most likely to be dirty glassware. You can purchase clean solutions that are sampled directly form the plastic delivery bottle. Also we do see many dirty stir bars - so please do not use a stirring bar unless you know it is absolutely clean.