SamplerSight : permutation entre types d'instrument

SamplerSight: Switching Instrument Type (76.6 KB)

Télécharger le fichier ci-dessus pour voir tous les tableaux et figures.
La version HTML ci-dessous donne une vue d'ensemble. Télécharger le fichier pdf vers voir tous les tableaux et figures.

General

In SamplerSight, users are only prompted for the Instrument Type on initial start up of the application. This is convenience of the user, for they can select the type of instrument to run and the application remembers that setting.

There are occasions where a user needs to switch the Instrument Type; this document provides the steps on how this is done.

For Versions 2.1 Service Pack 2 and up, we have added an easy way to switch instruments. There is a new shortcut in start menu (from Desktop go to Start/Programs/Particle Measuring) of "SamplerSight/Select Instrument" to select when needing to change instrument type.

For Version 2.1 (2.1.060306) SP2 Release and Up - Preferred Method

  1. Select Start/Programs/Particle Measuring
  2. Select "SamplerSight/Select Instrument" (see example of listing below). Note: This can be sent to the desktop as a shortcut by right clicking then selecting Send To/Desktop (create a shortcut)
  3. Application should automatically start up, select the new instrument type you with to switch to for "Please select an Instrument" prompt then click OK button

(see pdf for images)

For Older Systems - Preferred Method

  1. Select Start then Run
  2. Type (include the quotes), "C:\Program Files\ParticleMeasuring\SamplerSight\SamplerSight.exe" -select
  3. Application should automatically start up, select the new instrument type you wish to switch to for "Please select an instrument" prompt then click OK button

(see pdf for images)

For Older Systems - Only Use as a Last Resort

Warning: this way will cause their data configuration settings (under Configure Data) to be reset to the defaults.

  1. Shutdown application
  2. Right click on Start then select Explorer
  3. Navigate to the C:\Program Files\ParticleMeasuring\SamplerSight directory
  4. Locate and delete SamplerSight.mdb file
  5. Close Windows Explorer
  6. Restart application then select the appropriate instrument type

Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801.