Muestreo de partículas en el aire de 5,0 micrones (582.2 KB)

Véase la descarga del archivo arriba para tener cuadros completos e ilustraciones.
La versión HTML a continuación proporciona una reseña rápida. Baje el PDF con todas las tablas y figuras.

Summary

Particle Counter Lasair III

As the foremost worldwide experts in aerosol particle counting, Particle Measuring Systems recommends against using a 0.1 CFM unit to sample and monitor 5.0 micron particles. Instead, a 1.0 CFM unit, such as the LASAIR® III, provides needed sample to ensure the detection of undesired 5.0 micron particles.

Particle Loss in Transport

In training several thousand students at the Particle Measuring Systems' Particle College, instructors have consistently stressed the difficulties of sampling large particles; the inertia in such particles is so great that they cannot effectively travel through the sample tubing without bumping into and adhering to the tubing walls. Thus, they will present to the particle counter, not when they first appear, but only when something bumps against the tubing, jarring them loose.

The well-documented exception occurs when the sample tubing is evacuated at an unusually high flow rate (e.g., 2.0 to 3.0 CFM). Such rates normally are seen only in particle manifold systems, such as the Aerosol Manifold II. These systems can transport large particles for longer distances, but cannot provide continuous monitoring of the sampling probe of interest. Even with the use of such high flows, the 5.0+ micron particles suffer such high fallout rates that a maximum of 20% can traverse 125' of tubing (see Figure 2) - and this only if the tubing is absolutely straight, with none of the normal bends.

Design Recommendations

As a result, Particle Measuring Systems' research shows that the most reliable way to count large particles is to design the system as follows:

1) Sample the maximum quantity of air, thereby maximizing the chances of a 5.0 micron particle being present. This is best done by increasing the flow rate to 1.0 CFM, which provides 10 times the sample to analyze as a 0.1 CFM flow.

2) Minimize the sample tubing, thereby minimizing large particle fallout.

a) NO Tubing: Given laminar flow, the particles sampled will be those present in the cylinder of sample air that enters the sample inlet of the particle counter. When isokinetic sampling at a constant vertical flow rate, the area of the base of the cylinder will be directly proportional to the sample flow rate.

For example: For a vertical flow rate of 90 ft/min, the inlet area of an isokinetic probe will be:

For 1.0 CFM probe: 18.00 mm radius => area of 1032 mm^2.

For 0.1 CFM probe: 5.75 mm radius => area of 103 mm^2.

Thus, with 1.0 CFM, the area for catching large particles is ten times as large.

b) Minimal Tubing: If tubing cannot be totally eliminated, then the higher the sample velocity through the tubing, the fewer large particles will be lost. In comparison:

At 1.0 CFM, the sample velocity is approximately 3,350 ft./min.

At 0.1 CFM, the sample velocity is only 1,100 ft./min. and will lose a much higher percentage of the 5.0 micron particles.

Contact us if you need more information or have questions.

Lasair® is a registered trademark of Particle Measuring Systems, Inc.

Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801.

For more information go to Lasair II 550L 1.0 cfm monitor.

Documentación Contáctenos Quiénes somos Mapa del sitio Referencias

Sin medición no hay control
© 2002-2008 Particle Measuring Systems, Inc.

La información sobre este sitio web está sujeta a cambios sin aviso previo.

Web Media Engineering - website design Denver Colorado