FAQ - Contador de partículas de líquidos
Envíe una nueva pregunta- Pregunta
If we set the data as counts per ml (normalized count), then the will the reading be interpreted as counts/ml directly? or what is the difference between raw count and normalized count? - Respuesta
Regarding an APSS-200 system with a LiQuilaz E20P sensor (2 -125 um)
The data you view is based upon one criterion: do you manufacture Large or Small Volume Injectables (the definition would follow either USP, EP or JP standards)?
Small Volume (SVI) settings need to be: Accumulative, Normalized to Counts per Container.
This is, you need to see particles of size X and larger and this will encompass all particles of that size, and the legislation allows for a certain maximum particles in a single dose, and in SVIs this is deemed to be a unit volume of a container. Now USP/EP will require several (>10) of these containers to be pooled and the SamplerSight software allows for this in the configuration. The difference between Normalized and Raw is that normalized will allow for minor tolerance in a fixed volume, so you could take a 5 mL sample and the results will be essentially an average of the total counts / total volume = counts per unit volume. This is critical for compliance. Raw counts are the actual particles in suspension that are seen during a sample run.
Example:
Sample = 5mL, total counts = 125
Raw counts = 125
Counts / mL = 25
Large Volume (LVI) settings need to be: Accumulative, normalized to counts per mL.
Again, all particles of size x and greater and now due to the volume of dosage the criteria is based upon /mL as it is unknown how much final solution will be delivered.
The environment for both USP and EP needs to be in a laminar flow cabinet, EP define and ISO5 limit. SamplerSight has a calibration wizard included and this again should be followed with the instrument in a laminar flow cabinet.
For more information on understanding results from liquid particle counters I suggest you purchase the book Understanding Liquid Particle Counters.
- Soporte al cliente
- Soporte técnico
- Boletín informativo
- Preguntas frecuentes
- Partículas del aire
- Contador de partículas de líquidos
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Distribuciones de partículas
- Flow rate
- PSLs
- Período de estabilización
- Plumb
- Post Service
- Stable counts
- Statistically valid
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- LaminarCabinet
- Counts/mL Reading
- APSS 200 parameters
- Particles per mL
- Liquistat Nitrogen needed
- HSLIS Pressure
- Conteo de partículas en producción farmacéutica y ciencias de la vida
- Servicio de contadores de partículas
- Monitoreo molecular
- Análisis de gases
- General
- Documentos técnicos
- Contadores de partículas
- Monitores moleculares
- Servicios
- Alquiler y Arriendo
- Noticias y boletines de prensa
La información sobre este sitio web está sujeta a cambios sin aviso previo.