I have surfactant in my solution; will my particle counter mistake bubbles for particles? - Respuesta
One of the most important aspects to consider with surfactants is handling the product. Shaking, agitating, or even filling can entrain excessive air into the sample. All of our liquid particle counters will count the bubbles; in fact, any contaminant with a refractive index different from the solution will reflect light and therefore be recorded as a particle. Our compression samplers on the other hand, compress the sample prior to taking the reading to help reduce false counting bubbles as particles; for this reason, we would recommend these instruments for this application. Bear in mind, compression samplers can only minimize bubbles, not eliminate them completely especially artificially entrained air.
- Soporte al cliente
- Soporte técnico
- Boletín informativo
- Preguntas frecuentes
- Partículas del aire
- Contador de partículas de líquidos
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Distribuciones de partículas
- Flow rate
- PSLs
- Período de estabilización
- Plumb
- Post Service
- Stable counts
- Statistically valid
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- LaminarCabinet
- Counts/mL Reading
- APSS 200 parameters
- Particles per mL
- Liquistat Nitrogen needed
- HSLIS Pressure
- Conteo de partículas en producción farmacéutica y ciencias de la vida
- Servicio de contadores de partículas
- Monitoreo molecular
- Análisis de gases
- General
- Documentos técnicos
- Contadores de partículas
- Monitores moleculares
- Servicios
- Alquiler y Arriendo
- Noticias y boletines de prensa
La información sobre este sitio web está sujeta a cambios sin aviso previo.