I am seeing high DC light, what could be the problem? - Respuesta
DC light measures the amount of background scatter in some liquid instruments, mainly the LiQuilaz particle counter line. It is important to monitor the DC light level as excess background scatter can lead to a number of issues. High DC light usually occurs when the capillary is contaminated. In most cases, a contaminated capillary can be cleaned with most glassware cleaners. Samples with very high particle counts, bubbles, or high molecular weight additives can also lead to high DC light.
For more information read LiQuilaz Maintenance: Capillary Cleaning Procedure.
- Soporte al cliente
- Soporte técnico
- Boletín informativo
- Preguntas frecuentes
- Partículas del aire
- Contador de partículas de líquidos
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Distribuciones de partículas
- Flow rate
- PSLs
- Período de estabilización
- Plumb
- Post Service
- Stable counts
- Statistically valid
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- LaminarCabinet
- Counts/mL Reading
- APSS 200 parameters
- Particles per mL
- Liquistat Nitrogen needed
- HSLIS Pressure
- Conteo de partículas en producción farmacéutica y ciencias de la vida
- Servicio de contadores de partículas
- Monitoreo molecular
- Análisis de gases
- General
- Documentos técnicos
- Contadores de partículas
- Monitores moleculares
- Servicios
- Alquiler y Arriendo
- Noticias y boletines de prensa
© 2002-2008 Particle Measuring Systems, Inc.
La información sobre este sitio web está sujeta a cambios sin aviso previo.