Do you have any new information on foreign particle detection and limits for dry powder inhalers. I am current on the FDA's lack of momentum but looking to see if you could offer an insight on global work in this area or domestic work that I may not have seen. - 答案
Yes, unfortunately the extent of the FDA in helping assign limits is still 21 CFR 820 where a Quality Procedure should define the product/process. This is less than useful.
I have been involved in some studies looking at sterile closure cleanliness and in particular following the USP <381> test. This requires a cleanliness function that uses turbidity as the means to measure, however once on gets as far as turbidity you are not dealing with individual particles but a cloud of particles. Testing must therefore remain the same as we discussed last year, i.e. test sufficient of your own products to verify that an acceptable baseline exists (USP<1086> allows for certain impurities to exists providing they do not have an undesirable biological reaction) and then use this 95% confidence level as your limit for release.
Malvern Instruments (a sister company to Particle Measuring Systems, part of the Specris PLC Group) has a product called Spraytec that has particle size analysis for inhalants. If the testing above proves that residual particles of a certain size are present then this product would determine its function on overall particle distribution of finished, filled, delivered product.
That is the limit of business to date, unfortunately the emphasis is still on you the manufacturer to prove it is safe rather than have legislation drive the process.
Contact us if you need more information or have questions.
- 客户支持
- 技术支持
- 常见问题及答案
- 气体悬浮物粒子监测
- 液体粒子
- 制药和生命科学领域粒子监测
- Counter Height
- Room Qualification
- Monitoring Time
- ISO Hospitals
- CleanroomVerification
- ISO8
- Particle Count Frequency
- GMP testing
- IQ/OQ
- Support Docs
- Powder FMS
- Powder SOP
- Powder Collection Height
- Stopper Bowl
- Powder Regulations
- Filling Line
- Airnet Class 5
- Capping Station
- Cubic Meter Certification
- Isolated Excursions
- Sample point placement
- Dry Powder FMS
- IV Solutions
- Pharmaceutical Net
- EU GMP Annex I 2007
- 1986 Specifications
- 1986 v 1995 Specs
- 粒子监测服务
- 分子监测
- 气体分析
- 总类
- 新闻邮件
- 技术文章
- 粒子监测仪
- 活性粒子监测
- 分子监测仪
- 技术服务
- 仪器租赁
- 公司新闻
本网站所载信息未经通知既可能变化。