Wehn sampling a room to do a room qualification with the Lasair II, at what height should the samples be taken? Close to the floor or at working height? - 答案
ISO states that particle counting should be done at work height. In a room near a work place this is typically 1.0 m +/- 10%; in a corridor or walk way this is 1.5 m +/- 10%. Absolutely not on the floor.