I am trying to find copies of guidelines from the FDA or related documents about what the safe cutoffs are for IV solutions etc. I am an in vitro biochemist. I have been using a nephelometer to determine solubility of some compounds but I have some concerns about how accurate this approach will be. - 答案
There is one specific document that you need to read, USP <788>. The current version of USP is 30, it is released annually and is combined with the NF (United States Pharmacopoeia & National Formulary). Test 788 refers to particulate matter in injectables (this includes all injectables, SVP & LVP), the first part is that it must be free from all visible particles, these are deemed to be those that are above 100 um, and then set limits are imposed for sub visible particles, 10 & 25 um in size. We have an application note on our website that will point you in the right direction regarding numbers.
The above requires that a particle counter is used to enumerate the values here; however, if your product is an emulsion, suspension or similar then it can be tested using a microscope method. We have also seen issues where relatively viscous solutions (protein rich) also require filtration due to unsuitability issues.
Particle Measuring Systems offers a service where we can baseline your solutions using our laboratory here in Boulder.
Contact us if you need more information or have questions.
- 客户支持
- 技术支持
- 常见问题及答案
- 气体悬浮物粒子监测
- 液体粒子
- 制药和生命科学领域粒子监测
- Counter Height
- Room Qualification
- Monitoring Time
- ISO Hospitals
- CleanroomVerification
- ISO8
- Particle Count Frequency
- GMP testing
- IQ/OQ
- Support Docs
- Powder FMS
- Powder SOP
- Powder Collection Height
- Stopper Bowl
- Powder Regulations
- Filling Line
- Airnet Class 5
- Capping Station
- Cubic Meter Certification
- Isolated Excursions
- Sample point placement
- Dry Powder FMS
- IV Solutions
- Pharmaceutical Net
- EU GMP Annex I 2007
- 1986 Specifications
- 1986 v 1995 Specs
- 粒子监测服务
- 分子监测
- 气体分析
- 总类
- 新闻邮件
- 技术文章
- 粒子监测仪
- 活性粒子监测
- 分子监测仪
- 技术服务
- 仪器租赁
- 公司新闻
本网站所载信息未经通知既可能变化。