How do Airnet 510 particle sensors (1 cfm) comply to ISO 14644-1 standard for Class 5 (29 counts per cubic meter @ 5.0 micron)? Currently, we've worked out all sampling intervals to be 25 minutes, in order to achieve 690 Litres of sample air based on the ISO 14644-1 standard for Class 5.
Resposta

ISO14644-1 is the room certification; it is performed as per the exact standard issued. The first step for pharmaceutical customers is to certify the cleanroom to ISO standards. Once performed they now move to step 2 - cleanroom monitoring.

The big difference between the two is that room certification follows a formula and is used to calibrate the HVAC system. Monitoring is a risk based approach used to prove continued compliance of the room. The only point at which these two exercises ever meet is that in having an installed system the frequency of re-certification can be extended to 24 months (ISO14644-2).

You cannot certify a cleanroom using a monitoring system, only extend the interval between re-certifications. The sample points are in the wrong place etc.