Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute? - Resposta
All particle counters must have fluid flow at their rated calibration flow rate. Passing fluid through the particle counter at flow rates exceeding the calibrated flow will cause the instrument to report particle size and concentration inaccurately. There are currently no particle counters on the market that have a flow rate in excess of 1-liter/minute; however this does not limit the ability to measure in high volume applications. You can simply tee off the main line and connect this lower flow rate to the particle counter. It is usually necessary to provide some type of flow controller down stream of the particle counter to regulate the flow within the calibration limits of the sensor. You can then apply the normalized counts/ml value reported by the particle counter to the entire flow stream.
- Suporte ao consumidor
- Documentos Técnicos
- Newsletters
- PERGUNTAS FREQÜENTES (FAQ)
- Partículas aerotransportadas
- Partículas líquidas
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Distribuições de partículas
- Flow rate
- PSLs
- Período de estabilização
- Plumb
- Post Service
- Stable counts
- Statistically valid
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- HSLIS Pressure
- Contagem de partículas em laboratórios farmacêuticos ou de ciências biológicas
- Emprego dos contadores de partículas
- Monitoramento molecular
- Análise de gases
- Geral
- Calibração / Serviços
- Contadores de Partículas
- Amostrador de ar: monitoramento microbiológico
- Analisadores Moleculares
- Serviços
- Aluguel & Leasing
A Informação deste Site está sujeita à alterações a qualquer momento sem aviso.