I have surfactant in my solution; will my particle counter mistake bubbles for particles? - Resposta
One of the most important aspects to consider with surfactants is handling the product. Shaking, agitating, or even filling can entrain excessive air into the sample. All of our liquid particle counters will count the bubbles; in fact, any contaminant with a refractive index different from the solution will reflect light and therefore be recorded as a particle. Our compression samplers on the other hand, compress the sample prior to taking the reading to help reduce false counting bubbles as particles; for this reason, we would recommend these instruments for this application. Bear in mind, compression samplers can only minimize bubbles, not eliminate them completely especially artificially entrained air.
- Suporte ao consumidor
- Documentos Técnicos
- Newsletters
- PERGUNTAS FREQÜENTES (FAQ)
- Partículas aerotransportadas
- Partículas líquidas
- Is it possible to perform in line particle counting in a emulsion when passing through a pipe at 450 litres / minute?
- Bubbles
- CLS Height
- Distribuições de partículas
- Flow rate
- PSLs
- Período de estabilização
- Plumb
- Post Service
- Stable counts
- Statistically valid
- Surfactants
- Surfactant miscibiity
- Colored Chemical
- Cleaning
- Compression Sampler Settings
- DC Light
- Software
- HSLIS Pressure
- Contagem de partículas em laboratórios farmacêuticos ou de ciências biológicas
- Emprego dos contadores de partículas
- Monitoramento molecular
- Análise de gases
- Geral
- Calibração / Serviços
- Contadores de Partículas
- Amostrador de ar: monitoramento microbiológico
- Analisadores Moleculares
- Serviços
- Aluguel & Leasing
A Informação deste Site está sujeita à alterações a qualquer momento sem aviso.