My chemistry is viscous and colored; how will this affect the particle counts? - Resposta
Viscosity on its own does not affect particle counting. Other concerns should be maintaining the proper flow rate, being able to draw the sample into a sampler, not cavitating the solution, and minimizing entraining air into the sample. The color of a solution can affect counting. If the color absorbs the laser or scatters additional light, the counts can be affected.